I am trying to control my TDS400+ TDS500+ TDS600 and TDS700 series product using the RS232 port and it is not working. Why?

Question:

I am trying to control my TDS400+ TDS500+ TDS600 and TDS700 series product using the RS232 port and it is not working. Why?

Answer:

The RS232 port on the TDS400, TDS500, TDS600 and TDS700 series products is a hardcopy port only and is not capable of controlling the instrument.  Control of the instrument can only be accomplished through the GPIB port.

This FAQ Applies to:

Product: TDS794D, TDS510A, TDS520A, TDS520C, TDS540, TDS540A, TDS540B, TDS540C, TDS580C, TDS684A, TDS684B, TDS684C, TDS744, TDS420A, TDS460A, TDS420, TDS460, TDS410, TDS544A, TDS724A, TDS744A, TDS754A, TDS782A, TDS784A, TDS620, TDS620A, TDS640A, TDS644A

FAQ ID : 55761

View all FAQs »