DMM7510 in 4W RTD outputs a square wave signal that affects Fluke Calibrator during calibration.

The reason there is what appears to be a square wave output when the RTD temperature measurement is turned on is because offset compensation (OCOMP) is enabled by default.

Эти часто задаваемые вопросы относятся к
DMM7510

729751
Can two Model 2657A be placed in parallel to gain higher current?

The 2657A can be connected in parallel with another 2657A to double the current available.
Please make sure that the LO is tied to ground when you are connecting the 2657A in parallel to double the current available.

Эти часто задаваемые вопросы относятся к
2657A

729096
How to clear a function without having to recycle power on the 2600B series.

The easiest way to delete a function is to set it equal to nil, e.g. myfunc = nil. You should then call the collectgarbage() function to free up the memory. Garbage collection will actually occur automatically, but it can take up to a few seconds to process.


Эти часто задаваемые вопросы относятся к
2612B
729091
Where can I find the battery discharge script for Model 2450?

Answer: The battery discharge script is on the tek forum.

Here is the link.

https://forum.tek.com/viewtopic.php?f=66&t=140141

 


Эти часто задаваемые вопросы относятся к
2450
729071
Comparison of DDC and SCPI Command codes for 6517.

Look the attached word document for comparison of DDC(Device Dependent Commands) to SCPI commands for Model 6517A.


Эти часто задаваемые вопросы относятся к
6517B
729066
How do I synchronize the outputs of multiple Model 3390 Arbitrary Waveform Generators?

 


Эти часто задаваемые вопросы относятся к
3390
729041
TTR500A Series Custom Calibration Kit

You can use a custom calibration kit as long as you know the calibration standards.You can create a calibration kit in VectorVu by following the steps below:


Эти часто задаваемые вопросы относятся к
TTR503A, TTR506A
724046
TTR500A Calibration

Yes, calibration needs to be done on the TTR500A series before use. Before making measurements with a VNA, you calibrate the VNA to reduce errors that can affect measurements. When you calibrate a VNA, you perform a mathematical correction to the measurement results.
Эти часто задаваемые вопросы относятся к
TTR503A, TTR506A

724036
Does the TTR500A series have De-Embedding Capability

The TTR500A does not have built in de-embedding capability, however the user could save .SNP files and post process the data in order to de-embed the DUT.


Эти часто задаваемые вопросы относятся к
TTR503A, TTR506A
724031
Does the TTR500A series have port extension capability.

The TTR500A series does support port extensions. An online tutorial on how to perform the port extension is shown in the below link. 

How to Perform Port Extension


Эти часто задаваемые вопросы относятся к
TTR503A, TTR506A
724026